Long description 962 portable spectral photometers with 0 °/45 ° geometry

Interchangeable screens (Optionally): 7 mm, 4 mm, 15 mm.
between devices correspondence: 0.2 ΔEab.
Short term repeatability: 0.10 ΔEab on white ceramics
Normal observer: 2 ° and 10 °.

0 °/45 ° measuring geometry
excellent device correspondence
Project functions
Tallness, clear LCDDisplay with graphics function
Light, compact, portables device
Variable measurement with and without aim device
Opazität, colour strength and colour assortment
Robust construction method
Independently of net by accumulator company

962 is an inexpensive spectral photometer with 0 °/45 °–measuring geometry. It delivers fast and reliably exact colour information on various materials as for example colour, varnish, paper, plastic and textiles.

Measuring functions and indexes
962 misses absolute values and difference values of the in the following performed colour metrical systems. The values can be given on each of nine light kinds for 2 ° and 10 ° normal observer: CIE XYZ, CIE Yxy, CIELAB, Hunter, CIELCH, CMC and white degree, yellow degree to E313 98 ASTM and Metamerie index to German Institute for Standardization 6172

Special project functions
The X Rite spectral photometer 962 enables to the user to define different colour standards and to summarise to projects.

Passport / Fail function
962 stores up to 1024 colour standards with passport / Fail Toleranzen. The passport / Fail assessment is indicated on the display as well as by red or green LED and an acoustic signal (with excess of the tolerance).

Interchangeable measuring screens (optionally)
With few handles the user can choose between different measuring screens with 4 mm, 7 mm and 15 mm of measuring opening. This function gives a quick, efficient adaptation possibility to the size of the measuring object to the operator.

Opazität, colour strength and colour assortment
962 misses Opazität as well as three kinds of the colour strength, visually, chromatically or 3 area value. Also 555 colour assortments is possible with 962. This are the important identification numbers for the colour quality inspection by the processing of plastics, textiles and materials with varnished surfaces.

Ergonomics friendly to operator
In addition to the programmes integrated in the device which support the user with the measurement the device is also friendly to operator itself. 962 is compact, easy and with the wrist loop and the lateral shaped recessed grips comfortably to hold. The announcements are big and clear. The accumulator company guarantees flexibility by the application of the device.

Device correspondence
962 has an overriding device correspondence – in particular according to other colour controlling systems. 962 offers an excellent correspondence to X Rites 964 as well as 968 0 °/45 spectral photometers. All devices communicate with X Rites quality assurance sound Rezeptiersoftware at Windows level.

Optional accessories
*Battery recharger unit
*Exchange accumulator stack
*Different change screens
*Different application facilities and measuring accessories
*UV filter
*Software for the high class control
*colour formulation software
*Service and servicing contracts
*Software trainings
*Colour metrics trainings

SPECIFICATIONS:

MODELS
962 (7 mm)
962page (4 mm)
962 l (15 mm)

CHANGE OPTICS (optionally)
4 mm of measurement (6.5 mm of target window)
7 mm of measurement (9.0 mm of target window)
15 mm of measurement (21.5 mm of target window)

SOURCE OF LIGHT
Gas filled wolfram lamp

LIGHT KINDS
A, C, D50, D65, D75, F2, F7, F11 and F12

NORMAL OBSERVER
2 ° and 10 °

RECEIVER
Silicon photodiodes with more raised
Blue delicacy

SPECTRAL RANGE
400 nm 700 nm

SPECTRAL INTERVALS
10 nm measurement
10 nm issue

STORAGE CAPACITY
1024 standards with tolerance
2000 tests

MEASURING AREA
0 200% of reflexion

MEASURING TIME
Little one 2 seconds

DEVICE CORRESPONDENCE
CIE L*a*b* 0.20 DE*ab average ones
Divergence on 12
BCRA Tiles series II
Maximum 0.40 DE*ab on one
Tile

SHORT TERM REPEATABILITY
0.10 DE*ab on white ceramics
(Standard divergence)

LIFE SPAN OF THE LAMP
approx. 500,000 measurements

ELECTRICITY SUPPLY
Interchangeable accumulator stack (nickel
Metallhydrid); 7.2 V 1.250 mAh

NET PART
180 240 VAC, 50 60 hertz, 15 W max.

LOADING TIME
approx. 4 hours – 100% of capacity

MEASUREMENTS PER LOAD
1,000 measurements within
8 hours

OPERATING TEMPERATURE
10 ° C to 40 ° C
maximum 85% of relative ones
Air humidity (not
condensing)

SAFEKEEPING TEMPERATURE
20 ° C to 50 ° C
WEIGHT
1.1 kg

DIMENSIONS
10.9 cm (height; 8.4 cm (width;
19.6 cm (length)

PROVIDED ACCESSORIES
Kalibrierstandards, operating instructions
on CD, net part,
Transport pocket

OPTIONAL ACCESSORIES
Battery recharger unit
Exchange accumulator stack

Weight 1
Dimensions (LxWxH) ---

Location Germany
Region NRW


View this article on the vendor's web site.